These couplers are available with a coupling ratio of 50:50, 70:30, 90:10, or 99:1. They are bidirectional, allowing any port to be used as an input. Please click the info icons info below for full specifications and operational diagrams pertaining to each coupler. An individual test data sheet is provided with each coupler; wideband coupler data sheets include coupling data and performance graphs that extend outside the specified bandwidth, covering the wavelength range where the coupling ratio remains within the specified tolerance. Details of our coupler testing procedures are provided on the Coupler Verification tab, and sample data sheets for our 840 nm 2x2 SM couplers can be viewed here: wideband and boxed wideband. Our couplers have undergone extensive testing to ensure they meet or surpass Telcordia requirements.
Name | Model | Description | Parameter | Price |
---|
● Available for any center wavelength, band width, coupling ratio
● Wide operating wavelength range and excellent flatness
● Available for various fiber options
● High stability of SOP against temperature
*SOP: State of Polarization
Optical Specifications
Parameter | Unit | Value |
Center Wavelength | nm | 840 |
Operating Wavelength BandWidth | nm | ±40 |
Max. Excess Loss | dB | 0.2 |
Min. Retern Loss | dB | 50 (Typ. ≧ 55 dB) |
Min. Directivity | dB | 50 (Typ. ≧ 55 dB) |
Fiber Type | - | Single mode fiber (Standard: Corning HI780C) |
Operating Temperature | ℃ | -5 to +75 |
Storage Temperature | ℃ | -40 to +85 |
Coupling Ratio | Grade | Coupling Ratio Tolerance | Signal Path (1x2: P1→P3) (2x2: P1→P3, P2→P4) Insertion Loss | Tap Path (1x2: P1→P4) (2x2: P1→P4, P2→P3) Insertion Loss | ||||
50 % | A | ± 6.0 % | ≦ | 3.8 | dB | ≦ | 3.8 | dB |
P | ± 4.0 % | ≦ | 3.6 | dB | ≦ | 3.6 | dB | |
30 % | A | ± 4.5 % | ≦ | 2.0 | dB | ≦ | 6. 1 | dB |
P | ± 3.5 % | ≦ | 2.0 | dB | ≦ | 6.0 | dB | |
20 % | A | ± 4.0 % | ≦ | 1.4 | dB | ≦ | 8.2 | dB |
P | ± 3.0 % | ≦ | 1.3 | dB | ≦ | 7.9 | dB | |
10 % | A | ± 3.5 % | ≦ | 0.8 | dB | ≦ | 12.1 | dB |
P | ± 3.0 % | ≦ | 0.8 | dB | ≦ | 11.7 | dB | |
1 % | A | ± 0.6 % | ≦ | 0.3 | dB | ≦ | 24.2 | dB |
P | ± 0.5 % | ≦ | 0.3 | dB | ≦ | 23.2 | dB |
※Custom-designed parts available.
Typical Optical Specifications
How to read the specification sheet
Center Wavelength
This is the center wavelength when designing a optical fiber coupler. And it is the center wavelength when conducting the inspection. It is recommended to select the center wavelength of the optical fiber coupler according to the center wavelength of the light source used.
Operating Wavelength Band Width
If a light source with a wide wavelength band (such as an SLD light source or a swept source) is used, Select a coupler with a wide operating wavelength range. However, it is not recommended to select a coupler with a wider operating wavelength than is necessary for the bandwidth of the light source used. Optical fiber coupler with a wider operating wavelength range tend to have greater wavelength dependence. It is recommended that the operating wavelength range be determined appropriately according to the characteristics of the light source used.
Excess Loss: EL
The percentage of light lost in the coupler. The smaller the excess loss, the more efficient the power of the light source can be used. Excess loss at P1 input is calculated by the following formula.
EL(P1)(dB)= -10log((P3(mW)+ P4(mW))/P0(mW))
Insertion Loss: IL
ImageInsertion loss includes both branch loss and excess loss.
The optical path with low loss (optical path with high transmittance) is called Signal Path (straight port), and the optical path with high loss (optical path with low transmittance) is called Tap Path (cross port).
The insertion loss of the P1 → P3 optical path is calculated by the following formula.
IL(P1→P3)(dB) = P3(dBm) – P0(dBm)
Directivity:Dir
Directivity is a measure of how well the coupler isolates two opposite-traveling (forward and reverse) signals at the coupled port.The figure below shows how to measure the directivity at P1 incidence. The directivity for P1 input is calculated by the following formula.
Dir (P1→P3)(dB)= -10log((P2(mW)/P1(mW))
Return Loss:RL
Return loss is defined as the ratio of the output from that port to the port where the light source was input. The figure below shows an example of how to measure the return loss when P1 is incident. Return loss for P1 input is calculated by the following formula.
RL (P1)(dB) = (P1(dBm) - P1’(dBm)) - 3(dB)